DC MILLI-OHM METER
PROGRAMMER MANUAL
⎯ 28 ⎯
:CONFigure:TEMPerature(no query form)
Function:
Set the temperature function
Syntax:
:CONFigure:TEMPerature
Examples:
:CONFigure: TEMPerature sets the temperature function.
:CONFigure:TCOMpensate:RANGe <NR
2>(no query
form)
Function:
Set the TC function and range.
Syntax:
:CONFigure:TCOMpensate:RANGe <NR2>
Examples:
:CONFigure:TCOMpensate:RAMGe 0 sets the TC function and auto-
range.
:CONFigure: TCOMpensate:RAMGe 0.2 sets the TC function and 300m
ohm range.
<NR2> is in ohm unit.
DC MILLI-OHM METER
PROGRAMMER MANUAL
⎯ 29 ⎯
:CONFigure:TCOMpensate:COEFficient
Function:
Set or query temperature coefficient for TC function.
Syntax:
:CONFigure:TCOMpensate:COEFficient<NR1>
<NR1> is in the range from -9999 to 9999.
:CONFigure:TCOMpensate:COEFficient?
Returns:<String>
Examples:
:CONFigure: :TCOMpensate:COEFficient 3930 sets the temperature
coefficient to 3930ppm.
:CONFigure: :TCOMpensate:COEFficient? Return “+3930 ” to indicate
the temperature coefficient 3930ppm.
<NR2> is in ppm unit.
:CONFigure:TCOMpensate:TEMPerature
Function:
Set or query require temperature for the TC function.
Syntax:
:CONFigure:TCOMpensate:TEMPerature<NR2>
<NR2> is in the range from 0.00 to 100.0.
:CONFigure:TCOMpensate:TEMPerature?
Returns:<String>