GW Instek GDS-1000 Series Programming Manual User Manual

Page 11

Advertising
background image

COMMAND

DETAILS

21

:channel<X>:invert

<Boolean>

:channel<X>:invert?

:chan<X>:inv
<Boolean>
:chan<X>:inv?

Parameter

<X>

Channel

<NR1>

Channel invert

1/2 CH1/2

0 off

1 on

Example :channel1:invert

1

Inverts Channel 1

28B

:CHANnel<X>:MATH

Set

Query

Description

Selects or returns the math operation type.
Same as: Math key → F1

Syntax

< Long >

< Short >

:channel<X>:math

<NR1>

:channel<X>:math?

:chan<X>:math <NR1>
:chan<X>:math?

Parameter

<X>

Channel

<NR1>

Math operation

1/2

CH1 or CH2

0

Math off

1

Add

2

Subtract

3

FFT

Example1 :channel1:math

2

Channel 1 – Channel 2

Example2 :channel2:math

2

Channel 1 – Channel 2

Example3 :channel2:math

2

Runs FFT on Channel 2

29B

:CHANnel<X>:OFFSet

Set

Query

Description

Sets or returns the offset level for a channel. The
offset level range depends on the vertical scale.

GDS-1000 Programming Manual

22

Syntax

< Long >

< Short >

:channel<X>:offset

<NR3>

:channel<X>:offset?

:chan<X>:offs <NR3>
:chan<X>:offs?

Parameter

<X>

Channel <NR3> Offset level

1/2 CH1/2 ±0.5 –0.5V

~

+0.5V

(2mV/div~50mV/div)

±5.0 –5.0V

~

+5.0V

(100mV/div~500mV/div)

±50.0 –50.0V ~ +50.0V

(1V/div ~ 5V/div)

Example :channel1:scale

1.00e–2

:channel1:offset 2.00e–2

Sets the Channel 1
scale to 10mV/div
Sets the Channel 1
offset to 20mV

30B

:CHANnel<X>:PROBe

Set

Query

Description

Sets or returns the probe attenuation factor.
Same as: Channel key → F4

Syntax

< Long >

< Short >

:channel<X>:probe

<NR3>

:channel<X>:probe?

:chan<X>:prob <NR1>
:chan<X>:prob?

Parameter

<X> Channel <NR1> Probe attenuation factor

1/2 CH1/2 0 1x

1 10x

2 100x

Example :channel1:probe

1

Sets the Channel 1
probe attenuation
factor to 10x

Advertising