GW Instek GDS-1000 Series Programming Manual User Manual
Page 22

COMMAND
DETAILS
43
Example :ref1:display
1
Turns on the reference
waveform A
68B
:REF<X>:LOCate
Set
Query
Description
Moves or returns the position of a reference
waveform.
Same as: Save/Recall key → F5 → Variable knob
Syntax
< Long >
< Short >
:ref<x>:locate
<NR1>
:ref<x>:locate?
:ref<x>:loc <NR1>
:ref<x>:loc?
Parameter
<X>
Reference
<NR1>
Position
1 A
–100 to +100
2 B
Note
Before using this command, turn on a reference
waveform. See the example below.
Example :ref1:display
1
:ref1:locate 0
Turns on the reference
waveform A and move
it to ±0 position
69B
:REF<X>:SAVe
Set
Description
Saves an input signal as a reference waveform.
Same as: Save/Recall key (save) → F2 → F2 → F3
Syntax
< Long >
< Short >
:ref<x>:save <NR1>
:ref<x>sav <NR1>
Parameter
<X>
Reference
<NR1>
Source
1 A
1
Channel 1
2 B
2
Channel 2
3
Math
GDS-1000 Programming Manual
44
Example :ref1:save
1
Saves the Channel 1
signal as the reference
waveform A
70B
*SAV
Set
Description
Saves the current panel settings into the internal
memory.
Same as: Save/Recall key
→ F1
Syntax *sav
Parameter
<NR1>
Internal memory
1 to 15
S1 to S15
Example *sav
1
Saves the current panel
settings into S1