Probe compensation, Maintenance, Press the utility key – GW Instek GDS-2000 series Uer manual User Manual
Page 159: X1 x10, Utility

MAINTENANCE
159
8. The calibration for Channel1
starts and ends
automatically, in less than 5
minutes.
Ch1 calibration 1/3
9. When finished, connect the
calibration signal to
Channel2 and press F5.
Channel2 calibration starts.
Done!!
10. (for 4 Channel model only) Repeat the above
step for Channel 3 and 4.
11. When the calibration for all channels is
completed, the display goes back the default
state.
Probe Compensation
Panel operation
1. Connect the probe between Channel1 input and
the probe compensation output (2Vp-p, 1kHz
square wave) on the front panel. Set the probe
attenuation to x10.
Utility
Help
CH1
CH2
MATH
CH3
CH4
CH1
MENU
HORI
MENU
CH2
CH3
CH4
CAT
300V
M
16pF
MAX. 300Vpk
1
CAT
300V
M
16pF
MAX. 300Vpk
1
2V
POSITION
HORIZONTAL
TIME/DIV
LEVEL
TRIGGER
VARIABLE
STBY
ON/
POSITION
POSITION
POSITION
VERTICAL
POSITION
GDS-2104
Digital Storage Oscilloscope
100MHz 1G Sa/s
VOLTS/DIV
VOLTS/DIV
VOLTS/DIV
VOLTS/DIV
2V
X10
X1
CH1
x1
x10
2. Press the Utility key.
Utility