Edit: template (boundary) – GW Instek GDS-2000 series Uer manual User Manual
Page 72

GDS-2000 Series User Manual
72
Stop
The test stops when the NoGo
condition is met. The buzzer
does not sound.
Stop+
The test stops and the buzzer
sounds when the NoGo
condition is met.
Continue
The test continues even when
the NoGo condition is met. The
buzzer does not sound.
Continue+
The test continues even when
the NoGo condition is met. The
buzzer also sounds.
Note
If the sound is turned Off in the buzzer setting
(page70), the sound is not produced even when
selecting Stop/Continue+ .
Edit: Template (boundary)
Background
The NoGo template sets the upper and lower
amplitude boundary. Two methods are available:
Min/Max and Auto.
Min/Max
Selects the upper boundary (Max) and
lower boundary (Min) as separate
waveforms, from the internal memory.
Advantage: The template shape and the
distance (allowance) between the source
signal are fully customizable.
Disadvantage: The waveforms
(templates) have to be stored internally
prior to this selection.