HT instruments HT401 User Manual

Page 22

Advertising
background image

HT401

EN - 21

Continuity test

Range Resolution

Accuracy

Protection

against

overcharge

600

 0.1

(0.8rdg+5dgt)

1000VDC/ACrms

Buzzer activated for R < 30

 and deactivated for R > 100

Capacity

Range Resolution

Accuracy

Protection against

overcharge

1.000

F 0.001F

(1.2rdg+2dgt)

1000VDC/ACrms

10.00

F 0.01F

100.0

F 0.1F

1.000mF 0.001mF
10.00mF 0.01mF

Maximum time for reaching results: 0.7s C<1mF; 3s C> 1 mF

Temperature with type K probe

Range Resolution Accuracy

(*)

Protection

against

overcharge

-40.0

 400 °C

0.1°C

(1.0rdg+10dgt)

1000VDC/ACrms

-40.0

 752 °F

0.1°F

(1.0rdg+18dgt)

(*) The specified accuracy applies at the reference environmental temperature ±1°C. If temperature differs by more than
5 degrees, to obtain the declared accuracy, a stabilization time of 2 hours is required

6.1.1. Electrical

characteristics

Conversion: TRMS
Sampling frequency:

3 times per second

Temperature coefficient:

0.15x(accuracy) /°C, <18°C or >28°C


6.1.2. Reference

standards

The instrument complies with standards: IEC/EN 61010-1, UL61010-1,
Insulation:

double insulation

Pollution level:

2

Overvoltage category:

CAT IV 600V, CAT III 1000V

Max height of use:

2000m (6561ft)


6.1.3. General

characteristics

Mechanical characteristics
Size (with shell):

190(L) x 94(W) x 48 (H) mm (7.5 x 3.7x 1.9 in)

Weight (battery included):

460g (1Lv)

Power supply
Battery type:

1x9V battery NEDA1604, JIS006P, IEC6F22

Low battery indication:

symbol “ ” at display

Battery duration:

approx. 150 hours (no backlight)

Auto Power OFF:

after 20 minutes of idleness

Fuses:

FUSE71: F11A/1000V, 20kA

FUSE72: F440mA/1000V, 10kA

Display
Characteristics:

4 LCD, 6000 dots plus decimal sign, bargraph
and backlight

Over range indication:

O.L” or “-O.L