6 final installation, 7 restrictor removal and troll 9500 shipping, Instruction sheet – In-Situ S2XP Restrictor for TROLL 9500—Installation and Calibration Instructions User Manual

Page 4

Advertising
background image

Instruction Sheet

5. In Win-Situ 4, click Run in the Zero Oxygen Point

screen. Follow the on-screen instructions to complete
the 0% calibration point.

6. When readings have stabilized, you will have several

choices:

l

Select Next, then click Finish to finalize the calibration.

l

Select Back to repeat the oxygen saturation point
(100%) calibration.

l

Select Run to repeat the zero oxygen point (0%)
calibration. Follows steps 5-6 to repeat the calibration.

Final Installation

1. Remove the RDO sensor from the calibration solution.

Rinse the RDO sensor thoroughly.

2. Loop the RDO cable that extends from the restrictor

cap and hold it so that it abuts the RDO sensor. See
Figure 1.9.

Figure 1.9 Looping the RDO cable for installation

3. Insert the RDO sensor and cable loop into the

restrictor cap, using the hole for the sensor and the
slot for the cable.

4. Push the sensor in until it snaps into the restrictor cap.

Push the cable down into the slot until a small loop
remains. See Figure 1.10.

Figure 1.10 RDO sensor and cable fully installed

5. Grip restrictor B firmly. Attach restrictor A to the

restrictor cap. Do not allow restrictor B to rotate, as
the RDO cable can become twisted or damage the
other sensors. See Figure 1.11.

Figure 1.11 Installing restrictor A —do not rotate restrictor B

Restrictor Removal and TROLL 9500
Shipping

1. Grip restrictor B firmly. Remove restrictor A from the

restrictor cap. Do not allow restrictor B to rotate, as
the RDO cable can become twisted or damage the
other sensors. See Figure 1.12.

Figure 1.12 Removing restrictor A —do not rotate restrictor B

2. Remove the RDO sensor and cable loop from the

restrictor cap. The end of the RDO cable will still be
attached to the TROLL 9500 body. See Figure 1.13.

Figure 1.13 RDO sensor removed from the restrictor cap

3. Grip the TROLL 9500 body firmly. Remove restrictor

B, being careful not to twist the RDO cable.

4. Remove the pH/ORP sensor by inserting the removal

tool (located in the accessories kit) underneath the
sensor. Note the sensor's port location (port 1) for
reinstallation. See Figure 1.14.

800-446-7488

4

www.in-situ.com

Advertising
This manual is related to the following products: