Wavelength selection, Adjusting the concentration calculation factors, Selecting the unit of measurement – Jenway 6850 User Manual

Page 42

Advertising
background image

42

Absorbance Difference 2

Purity Ratio = Abs@260nm / Abs@230nm

RNA

Conc (μg/ml) = (Abs@

260

x 49.1)

– (Abs@

230

x 3.48)

Protein Conc (μg/ml) = (Abs@230 x 183) – (Abs@260 x

75.8)

The option to include the third wavelength as a reference is given after the measurement mode is

selected. Use the up and down arrow keys to select Yes or No and press the Enter key to confirm.

Selecting the Yes option will result in all the absorbance value measured at wavelength 3 being

subtracted from the photometric values measured at wavelengths 1 and 2.

9.3

WAVELENGTH SELECTION

The measurement wavelengths of the selected measurement mode can be adjusted by pressing the

GOTO

key on the instrument keypad. The required wavelengths can then be entered using the

numerical keypad. Press the Enter key to confirm.

9.4

ADJUSTING THE CONCENTRATION CALCULATION FACTORS

The measurement mode option can be selected by

pressing the function key below the Coeff option.

The new concentration calculation factors can be

entered using the numeric keypad. Press the Enter key

to confirm.

9.5

SELECTING THE UNIT OF MEASUREMENT

The unit of measurement can be selected by pressing the function key below the Unit option. The

required unit of measurement can then be selected from the available options of IU, mM/L, M/L,

μg/mL, mg/mL, mg/L, mEq, ppb, ppm, % and other. Press the Enter key to confirm.