Section 12, Maintenance and setvice, Routine maintenance – Jenway 6850 User Manual

Page 52: Lamp replacement, Tungsten lamp replacement, Lamp cover

Advertising
background image

52

SECTION 12

– MAINTENANCE AND SETVICE

12.1

ROUTINE MAINTENANCE

Ensure the external surfaces of the unit are clean and free from dust. The sample area should always

be kept clean and any accidental spillage should be wiped away immediately. To give added

protection when not in use, the unit should be disconnected from the mains supply and covered with

the optional dust cover.

12.2

LAMP REPLACEMENT

Warning! HOT.

Switch off the mains power supply and wait at least

20 minutes before attempting to replace the

tungsten or deuterium lamps.

12.2.1 Tungsten Lamp Replacement

1. Unscrew the 4 screws indicated (2 screws

on each side) and remove the cover of the

spectrophotometer.

.

2. Unscrew the 2 screws on the top of the

lamp chamber and remove the cover.

Lamp

Cover