Table 1-4. specifications, bit 4886, 4 lis – KEPCO BIT 4886 Operator Manual User Manual

Page 13

Advertising
background image

BIT 4886 OPR 120413

1-3

TABLE 1-4. SPECIFICATIONS, BIT 4886

SPECIFICATION

DESCRIPTION

Output Voltage (Main
Channel)

High Range Programs BOP to rated value

Low Range Programs BOP to 1/4 rated value

Rise and Fall times

(VOLT:MIN to VOLT:MAX)

<100 microseconds

Temperature Coefficient

Full scale:

± 35 ppm/°C max

Zero: High Range: ±10 ppm/°C max

Low Range:

±30 ppm/°C max

Voltage Readback:

High Range

± 20

µV/°C max

Low Range

± 10

µV/°C max

Current Readback:

±10 ppm/°C max

Optical Isolation

Digital and Analog grounds can be separated by a maximum of 500 Volts.

Digital Input Format

Byte-Serial/IEEE

Power Requirement

Supplied by host power supply

Resolution

Main Channel Programming 15 bits (16 bits including ± sign)

Limit Channel Programming 12 bits

Readback 15 bits (16 bits including ± sign)

Linearity

Programming ± 1 LSB

Readback ± 2 LSBs

Data Readback
Accuracy

Voltage 0.05% of Max. Voltage

Current 0.05% of Max. Current

List

Step Range 500 microseconds to 10 seconds

Step Accuracy –5 TO +20 microseconds (See Note 1.)

Number Of Steps 1002

Rise/fall Time >100 microseconds

Transient

Transient Range 500 microseconds to 10 seconds

Transient Accuracy –20 TO +1200 microseconds (See Note 1.)

Rise/fall Time >100 microseconds

Environmental
Specifications

Operating Temp. 0 to 50° C

Storage Temp. –20 to +85° C

NOTES: 1. Accuracy is affected by GPIB and RS-232 communications. To attain these accuracies only the Serial poll command of

the GPIB interface should be used while executing a transient or list.

Advertising