Step 18, Ps 16, Step 15 – KEPCO BIT 4886 Operator Manual User Manual

Page 34: Step 23

Advertising
background image

3-12

BIT 4886 120413

12.At the calibration panel (Figure 3-4) click CURR to begin current calibration of the unit. Con-

nect CALIBRATION RESISTOR! window opens.

13.Connect load (see Table 3-3) across the output of the unit and connect DVM across the load

as shown in Figure 3-2. Voltage measurements must be converted to Amperes using Ohm’s
law, E = IR. Click OK button to close Connect CALIBRATION RESISTOR window.

14.The test window reads

LCurr ZERO

and the message Adjust for ZERO R83 (Io Zero) on

BOP A1 card. will be displayed.

15.Adjust the R83 Potentiometer on A1 board of BOP, until the external DVM reads voltage

specified in Table 3-2, then click OK button.

16.The text window reads

LCurr MAX

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust output until DVM

reads voltage specified in Table 3-2 for Low Current Accuracy, then click OK button.

17.The text window reads

LCurr MIN

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM reads volt-

age specified in Table 3-2 for Low Current Accuracy, then click OK button. NOTE: If the
<<<— or —>>> button is depressed, return to step 16 after clicking OK.

18.The text window reads

Curr ZERO

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust the output until the

DVM reads the voltage specified in Table 3-2 for Main Current Zero, then click OK button.

19.The text window reads

Curr MAX

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust output until DVM

reads the voltage specified in Table 3-2 for High Current Accuracy, then click OK button.

20.The text window reads

Curr MIN

and the message Measure and Adjust. is displayed.

Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM reading is

as specified in Table 3-2 for High Current Accuracy. NOTE: If the <<<— or —>>> button is
depressed, return to step 19 after clicking OK.

21.The text window reads

Curr PROTECT

and the message Measure and Adjust. is dis-

played. Using the " and # and

!

and

"

buttons (see PAR. 3.3.2.1), adjust until DVM

reading is as specified in Table 3-2. Click OK when done.

22.Current calibration is complete. If desired, the current calibration can be repeated by clicking

Curr at the calibration panel (Figure 3-4).

23.To exit Calibration, click on the Calibration On button on the main panel (Figure 3-3). This

will open a panel that allows the calibration to be saved. Click SAVE to save the calibration
constants (both voltage and current), or CANCEL to return to the old calibration.