6 specifications, Table 1-2. series el specifications, Specifications -5 – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.22 through 3.41 User Manual

Page 15: Series el specifications -5

Advertising
background image

SERIES EL 062011

1-5

1.6

SPECIFICATIONS

Parameters unique to different models of Series EL loads are listed in Table 1-1. Specifications
common to all Series EL models are listed in Table 1-2.

TABLE 1-2. SERIES EL SPECIFICATIONS

SPECIFICATION

RATING/DESCRIPTION

CONDITION/COMMENT

MODE CHARACTERISTICS

Linearity vs. programming command

See Table 1-3

5 – 100%; accuracy is % of Full Scale value(2)

Regulation(1)

See Table 1-3

5 – 100%; regulation is % of Full Scale value(2)

Resolution (via computer control)

14 Bits

5 – 100% of Full Scale

Current readback
(Current mode)

Computer Accuracy:

±0.25%

5 – 100%; accuracy is % of Full Scale value(2)

Computer Resolution:

±15 Bits

5 – 100% of Full Scale

Parameter readback
(Current and Voltage)

Accuracy: ±0.25% (±1% for 50V models) Accuracy is % of Full Scale value(2)

Resolution:

±15 Bits

Temperature

coefficient:

100 ppm/°C

CONTROL CHARACTERISTICS

USB Interface

Uses IEEE 488.2 and SCPI

commands and queries.

Requires driver.

RS 232 Interface

Uses IEEE 488.2 and SCPI

commands and queries.

Baud rate: 38400
Parity: None
Data Bits: 8
Stop Bits: 1
Flow Control: None

GPIB Interface

Uses IEEE 488.2 and SCPI

commands and queries.

GPIB address 10 = default

Ethernet Interface

Uses IEEE 488.2 and SCPI

commands and queries.

Optional

GENERAL CHARACTERISTICS

a-c Line Power Input

120 VAC ±10%, 50/60 Hz or

240 VAC ±10%, 50/60 Hz

120V is factory default; 240V requires internal
configuration (see PAR 2.3.1)

Cooling

Forced Air (400 CFM)

Variable speed fans, cooling determined by load
demand.

Operating Tempera-
ture

0°C to 40°C

Dimensions (load)

19”W x 7”H x 24.5”D

See Figure 2-10.

Weight (load)

88 lbs.

Storage Requirements

–20 to +60°C, 15 to 80% RH

Relatively dust free environment

NOTE:
(1) Regulation specified after 15 minutes of operation at set power level.
(2) Accuracy and regulation are valid from 5% to 100% of the rated value. The accuracy (e.g., ±0.25% of a constant current setpoint)
is ±0.25% of the full scale value and not the setpoint.

Advertising