TABLE 2-2. REAR PANEL CONNECTIONS

FIG. 2-3 or 2-4

INDEX NO.

NAME

DESCRIPTION

1

120VAC 50/60HZ/
250VAC 50/60HZ/

POWER: IEC 60320-C13 line cord connector to connect
mains power (either 120 VAC 50/60 Hz or optional 240V AC
50/60 Hz). NOTE: changing mains power voltage requires
internal connector changes (see PAR. 2.3.1).

2

FUSE

FUSE: 120 V, 3AG, 3 Ampere Slo Blow (or 240V, 3AG, 1.5
Ampere Slo Blow)

3

LOAD (–)

Bus bar

UUT Connections. (–) Negative and (+) Positive: High cur-
rent connections to UUT. Connections to ±LOAD are typically
made using the supplied 3/8” low-resistance hardware (2
sets: bolt, nut and lockwasher).

4

LOAD (+)

Bus bar

5

LOAD (–)

terminal

(–) Negative Load Voltage Monitor. (CAUTION: Do not
attempt to pass load current through this connector—it is
only for UUT voltage measurement)

6

RS (–)

terminal

Remote Sense (voltage) Negative Connection (CAUTION:
Do not attempt to pass load current through this connector—
it is only for UUT sense connection)

7

RS (+)

terminal

Remote Sense (voltage) Positive Connection (CAUTION: Do
not attempt to pass load current through this connector—it is
only for UUT sense connection)

8

LOAD (+)

terminal

(+) Positive Load Voltage Monitor. (CAUTION: Do not
attempt to pass load current through this connector—it is
only for UUT voltage measurement)

9

ANALOG PGM

Isolated BNC

Connector

A 0 to 10V analog programming signal can alter the digital
setpoint of the CONTROL parameter (depending on mode
selection and digital setpoint) from 0 to full scale. See PAR.
3.15 for details and an example.

10

ANALOG I LOAD

Isolated BNC

Connector

This signal is 0 to 10V d-c corresponding to 0A to the config-
ured/calibrated full scale current. Output at center pin when
the load is drawing current from the UUT.

11

LAN

Not shown.

12

MSTR

Master Out

connector

Standard RS-522 connector. Connects from Master Out to
Slave In (SL-I) for multiple units in parallel (see Figure 2-9).
See PAR. 3.16 for master/slave configurations. See Table 1-
5 to
order RJ 22 master/slave cable.

13

RS-232

connector

Standard RS-232. Connects with a remote device using RS
232 protocols and SCPI commands (see Figure 2-3).

14

USB

connector

Connection to remote device for command, monitoring or
calibration. Use USB connector of master to communicate
with multiple loads in a master/slave configuration. (see Fig-
ure 2-7).

15

SL-O

Slave Out
connector

Standard RS-522 connector. Connects Slave Out to SL-I
(Slave In) of next slave if unit is a slave in a parallel configu-
ration using more than two units. Only use Kepco cable (see
Table 1-5) for master/slave connections (Figure 2-9).

16

SL-I

Slave In

connector

Standard RS-522 connector. Connects Slave Out to SL-I
(Slave In) of next slave if unit is a slave in a parallel configu-
ration using more than two units. Only use Kepco cable (see
Table 1-5) for master/slave connections (Figure 2-9).

17

GPIB

IEEE 488 connector

Optional. When installed, connects user GPIB card with EL
load using standard GPIB cable (not supplied).