B.39 [source:]current:protection[:level] command, B.40 [source:]current:protection[:level]? query, B.41 [source:]current:protection:state command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.22 through 3.41 User Manual
Page 67: B.42 [source:]current:protection:state? query, B.43 [source:]mode command, B.44 [source:]mode? query, B.39, Source, B.40, B.41

SERIES EL 062011
B-7
B.39 [SOURce:]CURRent:PROTection[:LEVel] COMMAND
CURR:PROT
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV] <NR2>
Long Form: [SOURce:]CURRent:PROTection[:LEVel] <NR2>
where: <NR2> = current protection setpoint
Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.
B.40 [SOURce:]CURRent:PROTection[:LEVel]? QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns: <NR2> = current protection setpoint
Description: Indicates current protection setpoint.
B.41 [SOURce:]CURRent:PROTection:STATe COMMAND
CURR:PROT:STAT
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT <bool>
Long Form: [SOURce:]CURRent:PROTection:STATe <bool>
where <bool> = 1 (enabled) or 0 (disabled)
Description: Enables (1) or disables (0) the current protection level set using CURR:PROT. This command
saves having to change the protection level to some value outside of the operating range.
B.42 [SOURce:]CURRent:PROTection:STATe? QUERY
CURR:PROT:STAT?
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: <bool> = 1 (enabled) or 0 (disabled)
Description: Indicates whether the current protection level set using CURR:PROT is enabled (1) or disabled
(0).
B.43 [SOURce:]MODE COMMAND
MODE
Syntax:
Short Form: [SOUR:]MODE <txt>
Long Form: [SOURce:]MODE <txt>
where: <txt> = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.
Description: Sets operating mode: When the operating mode is changed, the input is set to off. Send INPut ON to
apply the new mode setting to the load.
B.44 [SOURce:]MODE? QUERY
MODE?
Syntax:
Short Form: [SOUR:]MODE?
Long Form: [SOURce:]MODE?
Returns: <txt> = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.
Description: Indicates operating mode.
B.45 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] COMMAND
POW
Syntax:
Short Form: [SOUR:]POW[:LEV][:IMM][:AMP] <NR2>
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] <NR2>
Where <NR2> = power setpoint
Description: Establishes power setpoint. The power setpoint only affects the load when the load is in Power
mode of operation.
B.46 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]? QUERY
POW?
Syntax:
Short Form: [SOUR:]POW[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]?
Returns <NR2> = power setpoint
Description: Indicates power setpoint.