Installation – KROHNE OPTISONIC 6300 EN User Manual
Page 18

3
INSTALLATION
18
OPTISONIC 6300
www.krohne.com
07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en
3.5.1 Inlet, outlet and recommended mounting area
3.5.2 Long horizontal pipes
• Install on slightly ascending pipe section.
• If not possible, ensure adequate velocity to prevent air, gas or vapor from collecting in upper
part.
• In partially filled pipes, the clamp-on flowmeter will report incorrect flow rates, or not
measure.
Figure 3-1: Inlet, outlet and recommended mounting area
1 Min. 10 DN
2 Min. 5 DN
3 OK, 120°
CAUTION!
Especially for XT (eXtended Temperature) versions:
•
Always install the sensor at a non-insulated part of the pipe. Remove any insulation if
necessary!
•
Bend radius of cable plus connection box needs 10 cm / 4" additional non insulated pipe
section.
•
Always wear protections gloves.
Figure 3-2: Long horizontal pipes
.book Page 18 Tuesday, August 18, 2009 4:51 PM