ELECTRICAL CONNECTIONS

46

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

4.6.2 Modular inputs/outputs and bus systems

Current output active (only current output terminals C/C- have HART

®

capability),

modular I/Os

• U

int, nom

= 24 VDC

• I ≤ 22 mA
• R

L

≤ 1 kΩ

• X designates the connection terminals A, B or C, depending on the version of the signal

converter.

Current output passive (only current output terminals C/C- have HART

®

capability),

modular I/Os

• U

ext

≤ 32 VDC

• I ≤ 22 mA
• U

0

≥ 1.8 V

• R

L

≤ (U

ext

- U

0

) / I

max

• X designates the connection terminals A, B or C, depending on the version of the signal

converter.

CAUTION!
Observe connection polarity.

INFORMATION!
•

For the electrical connection of the bus systems, please refer to the separate documentation
for the respective bus systems.

Figure 4-18: Current output active I

a

Figure 4-19: Current output passive I

p