INSTALLATION

3

25

OPTIWAVE 7300 C

www.krohne.com

03/2014 - 4000172405 - HB OPTIWAVE 7300 R07 en

3.6.2 Standpipes (stilling wells and bypass chambers)

Use a standpipe if:

• There is highly conductive foam in the tank.
• The liquid is very turbulent or agitated.
• There are too many other objects in the tank.
• The device is measuring a liquid (petro-chemicals) in a tank with a floating roof.
• The device is installed in a horizontal cylindrical tank (refer to the end of this section)

Figure 3-9: Installation recommendations for standpipes (stilling wells and bypass chambers)

1 A stilling well solution
2 A bypass chamber solution
3 Air circulation hole
4 Level of the liquid

CAUTION!

Installation requirements

•

The standpipe must be electrically conductive.

•

The inside diameter of the standpipe must not be more than 5 mm / 0.2

¨

over the diameter of

the antenna (for a high-dielectric constant liquid).

•

The standpipe must be straight. There must be no sudden changes in internal diameter
greater than 1 mm / 0.04

¨

.

•

The standpipe must be vertical.

•

Recommended surface roughness: <

±

0.1 mm / 0.004

¨

.

•

Stilling well only: The bottom of the stilling well must be open.

•

Make sure that there are no deposits at the bottom of the standpipe.

•

Make sure that there is liquid in the standpipe.