Installation, 7 how to install the device on the tank – KROHNE OPTIWAVE 7300C EN User Manual
Page 28

3
INSTALLATION
28
OPTIWAVE 7300 C
www.krohne.com
03/2014 - 4000172405 - HB OPTIWAVE 7300 R07 en
Bypass chambers
Installation next to tanks containing one liquid and foam
• The top process connection of the bypass chamber must be above the maximum level of
liquid.
• The bottom process connection of the bypass chamber must be below the lowest measured
level of liquid.
Installation next to tanks containing more than one liquid
• The top process connection of the bypass chamber must be above the maximum level of
liquid.
• The bottom process connection of the bypass chamber must be below the lowest measured
level of liquid.
• Additional process connections are necessary for the liquids to circulate freely along the
length of the bypass chamber.
3.7 How to install the device on the tank
3.7.1 How to install a device with a flange connection
Equipment needed:
• Device
• Gasket (not supplied)
• Nuts and bolts (not supplied)
• Wrench (not supplied)
Figure 3-12: Installation recommendations for bypass chambers that contain more than one liquid
1 Bypass chamber
2 Additional process connection