9 supplement, 1 technical data – KROHNE OPTISOUND 3010C 2wire EN User Manual
Page 38

9 Supplement
9.1 Technical data
General data
Materials, wetted parts
- Process fitting
Thread G1½A and 1½ NPT: PVDF
- Transducer
PVDF
- Seal transducer/process fitting
EPDM
Materials, non-wetted parts
- Housing
plastic PBT (Polyester), Alu-die casting pow-
der-coated, 316L (1.4435)
- Seal ring between housing and
housing cover
NBR (stainless steel housing), silicone (Alu/
plastic housing)
- Inspection window in housing cover
Polycarbonate
- Ground terminal
316Ti/316L (1.4571/1.4435)
Weight
1.8 … 4.0 kg (4.0 … 8.8 lbs), depending on
process fitting and housing
Max. torque mounting boss
25 Nm
Output variable
Output signal
4 … 20 mA/HART
Resolution
1.6 µA
Fault signal
current output unchanged; 20.5 mA; 22 mA;
<3.6 mA (adjustable)
Current limitation
22 mA
Load
see load diagram in Power supply
Integration time (63 % of the input
variable)
0 … 999 s, adjustable
Fulfilled NAMUR recommendation
NE 43
Input variable
Parameter
distance between lower edge of the transducer
and product surface
Dead zone
0.25 m (0.8 ft)
Measuring range
- Liquids
up to 5 m (16.4 ft)
- Solid
up to 2 m (6.6 ft)
38
OPTISOUND 3010 C - 4 … 20 mA/HART
Supplement
30503
-EN
-050914