Permissible residual ripple - <100 hz u, 10 mv load see diagram – KROHNE OPTISOUND 3010C 2wire EN User Manual
Page 41
Advertising

Permissible residual ripple
- <100 Hz
U
ss
<1 V
- 100 Hz … 10 kHz
U
ss
<10 mV
Load
see diagram
1000
750
500
250
14
18
16
20
22
24
26
28
30
32
34
36
Ω
V
4
1
2
3
Fig. 26: Voltage diagram
1
HART load
2
Voltage limit EEx ia instrument
3
Voltage limit non-Ex instrument
4
Supply voltage
Electrical protective measures
Protection
- Plastic housing
IP 66/IP 67
- Alu and stainless steel housing
IP 66/IP 68 (0.2 bar)
5)
Overvoltage category
III
Protection class
II
Approvals
6)7)
ATEX
ATEX II 1G, 1/2G, 2G EEx ia IIC T6
5)
Requirement to maintain the protection is the suitable cable.
6)
Deviating data with Ex applications: see separate safety instructions.
7)
Depending on order specification.
OPTISOUND 3010 C - 4 … 20 mA/HART
41
Supplement
30503
-EN
-050914
Advertising