TECHNICAL DATA

28

OPTITEMP TR/TC 100

www.krohne.com

10/2009 - 4000630002 - MA OPTITEMP TR/TC 100 R02 en

7.3.2 Measuring insert with type J thermocouples

Type J: basic values and tolerances

Temperature

(T90) in °C / °F

Basic value

in mV

Thermal EMF

in mV/K

Tolerances acc. to DIN EN 60584-2

Class 1

Class 2

mV

°

C / °F

mV

°

C / °F

-40 / -40

-1.961

0.046

±

0.07

±

1.5 / 2.7

±

0.12

±

2.5 / 4.5

0 / +32

0

0.053

±

0.08

±

1.5 / 2.7

±

0.13

±

2.5 / 4.5

+100 / +212

5.269

0.055

±

0.08

±

1.5 / 2.7

±

0.14

±

2.5 / 4.5

+200 / +392

10.779

0.055

±

0.08

±

1.5 / 2.7

±

0.14

±

2.5 / 4.5

+300 / +572

16.327

0.055

±

0.08

±

1.5 / 2.7

±

0.14

±

2.5 / 4.5

+400 / +752

21.848

0.055

±

0.09

±

1.6 / 2.9

±

0.17

±

3.0 / 5.4

+500 / +932

27.393

0.057

±

0.11

±

2.0 / 3.6

±

0.21

±

3.8 / 6.8

+600 / +1112

33.102

0.056

±

0.13

±

2.4 / 4.3

±

0.25

±

4.5 / 8.1

+700 / +1292

39.132

0.064

±

0.18

±

2.8 / 5.0

±

0.34

±

5.3 / 9.5

+750 / +1382

42.281

0.064

±

0.19

±

3.0 / 5.4

±

0.36

±

5.6 / 10.1

Figure 7-4: Type J tolerances

1 Tolerances in °C
2 Tolerances in °F
3 Tolerance class 2
4 Tolerance class 1
5 Temperature in °C
6 Temperature in °F