3 instrument repair – KROHNE SU 600 EN User Manual
Page 28

?
E016
l
Empty/full adjustment reversed
à
Carry out a fresh adjustment
?
E017
l
Adjustment span too small
à
Carry out a fresh adjustment and increase the
distance between min. and max. adjustment
?
E021
l
Scaling span too small
à
Carry out a fresh scaling, increase the distance
between min. and max. scaling.
?
E110
l
Relay switching points too close together
à
Increase the difference between the two relay
switching points
7.3 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form from our Internet
homepage http://www.krohne-mar.com/fileadmin/
media-lounge/PDF-Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and
without having to call back for needed information.
l
Print and fill out one form per instrument
l
Clean the instrument and pack it damage-proof
l
Attach the completed form and possibly also a safety
data sheet to the instrument.
28
SU 600 - 4 … 20 mA signal conditioning
Maintenance and fault rectification