4/2

Condensed instructions IFC 020

π

4

π

4

4.2 Table of settable functions

Fct.

Text

Description and settings

1.00

OPERATION

Operations menu

1.01

FULL SCALE

Full-scale range for flowrate Q

100%

Select unit
• m

3

/hr

• Liter/Sec

• US.Gal/min

• user unit, factory set is “Liter/hr” or “US MGal/day” (see Fct. 3.05)

Press

→ key to transfer to number setting.

Setting ranges
The ranges are dependent on the meter size (DN) and the flow

velocity (v): Q

min

= DN

2

x v

min

Q

max

= DN

2

x v

max

Nom. dia./meter size

v

min

= 0,3 m/s (1 ft/s) v

max

= 12 m/s (40 ft/s)

• DN 2.5–1000 /

1

/

10

”–40”:

0.0053 –

0

33 900 m

3

/hr

0.0237 –

152 000 US.Gal/min

Press

↵ key to return to Fct. 1.01 FULL SCALE.

→ VALUE P

Pulse value (Fct. 1.06 “VALUE P”) has been changed.

With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

1.02

TIMECONST.

Time constant
Select:

• ALL (applies to display and all outputs)
• ONLY I+S (only display, current and status outputs)

Press

↵ key to transfer to number setting.

Range:

• 0.2 – 99.9 Sec

Press

↵ key to return to Fct. 1.02 TIMECONST.

1.03

L.F.CUTOFF

Low-flow cutoff (SMU)
• OFF (fixed values: ON = 0.1% / OFF = 0.2%

at 100Hz and 1000Hz, see Fct. 106, 1% resp. 2%)

• PERCENT (variable values)

ON OFF
1 – 19% 2 – 20%

Press

→ key to transfer to number setting.

Note: Cutoff “off” value must be greater than cutoff “on” value.
Press

↵ key to return to Fct. 1.03 L.F. CUTOFF.

1.04

DISPLAY

Display functions

→ DISP.FLOW

Select flow display
•
NO DISP. • user unit, factory set is “Liter/hr” or “US MGal/day (see Fct. 3.05)
• m3/hr

• PERCENT

• Liter/Sec

• BARGRAPH (value and bargraph display in %)

• US.Gal/min

Press

↵ key to transfer to subfunction “DISP. TOTAL.”.

→ DISP.TOTAL.

Select totalizer display
•
NO DISP. (totalizer switched on but not displayed)
• OFF (totalizer switched off)
•

+TOTAL

•

−TOTAL

•

±TOTAL

• SUM (

Σ)

• ALL (single totalizer or all)
Press

↵ key to transfer to display unit

• m3

• Liter

• US.Gal

• user unit, factory set is “Liter” or “US MGal” (see Fct. 3.05).

Press

→ key to transfer to format setting.

Format setting
• Auto (exponent notation)
• # . #######

• ##### . ###

• ## . ######

• ###### . ##

• ### . #####

• ####### . #

• #### . ####

• ########

Press

↵ key to transfer to subfunction “DISP.MSG”.

→ DISP.MSG.

Additional messages required in measuring mode?
•
NO

• YES (cyclic change with displays of measured values)

Press

↵ key to return to Fct. 1.04 DISPLAY.