KROHNE IFC 020 Short EN User Manual

Page 17

Advertising
background image

Condensed instructions IFC 020

4/3

Fct.

Text

Description and settings

1.05

CURRENT I

Current output I

→ FUNCT. I

Select function for current output I
•
OFF (switched off)
• 1 DIR. (1 flow direction)
• 2 DIR. (forward/reverse flow, F/R flow measurement)
Press key

↵ , transfer to subfunction “RANGE I”;

→ RANGE I

Select measuring range
•
0 - 20 mA • 4 - 20 mA (fixed ranges)
• mA (user-defined range) I

0%

- I

100%

(Value I

0%

< I

100%

!) 0 - 16 mA 4 - 20 mA

To transfer to number setting, press key

→ !

Press key

↵ to transfer to subfunction “I ERROR“.

→ I ERROR

Select error value
•
22 mA • 0.0 to I

0%

mA (variable, see above if I

0%

> 1 mA)

To transfer to number setting, press key

→ !

Press key

↵ to revert to Fct. 1.05 CURRENT. I.

1.06

PULS B1

Pulse output B1 (see Fct. 3.07 HARDWARE)

→ FUNCT. P

Select function for pulse output P
•
OFF (switched off)
• 1 DIR. (1 flow direction)
• 2 DIR. (forward/reverse flow, F/R flow measurement)
Press key

↵ to transfer to subfunction “SELECT P”.

→ SELECT P

Select pulse type
•
100 HZ

• PULSE/VOL. (pulses per unit volume, flowrate)

• 1000 HZ

• PULSE/TIME (pulses per unit time for 100% flowrate)

Press key

↵ to transfer to subfunction “PULSWIDTH”.

With selection 100 HZ and 1000 HZ,
return to Fct.1.06 pulsoutput P (pulswidth 50% cyclic).

→ PULSWIDTH

Select pulse width
•
50 msec • 100 msec • 200 msec • 500 msec • 1 sec
Press key

↵ to transfer to subfunction “VALUE P”.

→ VALUE P

Set pulse value per unit volume (appears only when
“PULSE/VOL.” set under “SELECT P” above)
• xxxx PulS/m3 • xxxx PulS/Liter • xxxx PulS/US.Gal
• xxxx PulS/ user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width and the
full-scale range: P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Press key

↵ to return to Fct.1.06 “PULS. OUTPUT P”.

→ VALUE P

Set pulse value per unit time (appears only when “PULSE/TIME”
set under “SELECT P” above)
• xxxx PulSe/Sec (=Hz) • xxxx PulSe/min • xxxx PulSe/hr
• xxxx PulSe/user-defined unit, factory-set is “hr” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width, see above
Press key

↵ to return to Fct.1.06 “PULS. OUTPUT P””.

1.07

STATUS. S

Status output S

• ALL ERROR

• FATAL ERROR

• OFF

• ON

• F/R INDIC. (F/R indication for forward/reverse measurement)
• TRIP. POINT

Setting range: 002 - 115 PERCENT

• EMPTY PIPE (appears only when this option is installed)
Press

↵ key to transfer to number setting

Press

↵ key to return to Fct. 1.07 “STATUS. S“.

1.08

CONTROL E

Control input E (IFC 020 E only)

(IFC 020 E only)

• OFF (switched off)
• OUTP. ZERO (set outputs to “min.values”)
• TOTAL. RESET (reset totalizers)
• ERROR. RESET(delete error messages)
Press

↵ key to return Fct. 1.08 “Control E”.