Fct.

Text

Description and settings

3.03

ZERO SET

Zero calibration
Note:

Carry out only at “0” flow and with completely filled
measuring tube!

Precautionary query
• CALIB. NO

Press

↵ key to return to Fct. 3.03 “ZERO SET”.

• CALIB. YES Press

↵ key to start calibration.

Duration approx. 15-90 seconds, (depends on frequency
of magnetic field) current flowrate displayed in the selected
unit (see Fct. 1.04 “DISP. FLOW”).

A “WARNING” sign appears when flowrate “> 0“;
acknowledge by pressing

↵ key.

• STORE NO (do

not store new zero value)

• STORE YES (store new zero value)
Press

↵ key to return to Fct. 3.03 “ZERO SET”.

3.04

ENTRY CODE

Entry code required to enter setting mode?
•
NO

(= entry with

→ only)

• YES (= entry with

→ and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑ )

Press

↵ to return to Fct. 3.04 “ENTRY CODE”.

3.05

USER UNIT

Set any required unit for flowrate and counting

→ TEXT VOL.

Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press

↵ key to transfer to subfunction “FACT. VOL.”

→ FACT. VOL.

Set conversion factor (F

M

) for volume

Factory set “1.00000 E+3” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1

x

10

3

or 2.64172

x

10

-4

).

Factor F

M

= volume per 1m

3

.

Setting range

• 1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

↵ key to transfer to subfunction “TEXT TIME”.

→ TEXT TIME

Set text for required time unit (max. 3 characters)
Factory-set: “hr” (=hour).
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press

↵ key to transfer to subfunction “FACT. TIME”

→ FACT. TIME

Set conversion factor (F

T

) for time

Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6

x

10

3

or 8.64

x

10

-4

).

Set factor F

T

in seconds.

Setting range
• 1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

↵ key to return to Fct. 3.05 “USER UNIT”.

3.06

APPLICAT.

Set overload point for A/D converter

→ FLOW

• STEADY (150% of Q

100%

) • PULSATING (1000% of Q

100%

)

Press key

↵ to return to Fct. 3.06 “APPLICAT.“,

3.07

MEASURING POINT

Set measuring point no.
Factory setting: ALTOMETER
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press

↵ key to return to Fct. 3.07”MEASURING POINT”

3.08

COM

Set comminucation interface
•
OFF (switched off)
• HART (HART-interface switched on)
• KROHNE (RS 485 interface switched on)
• Address:”HART” 00-15 / “KROHNE” 000-239
• BAUD RATE: -1200

-2400

-4800

-9600

-19200

(appeare with selection “KROHNE”)

Press key

↵ to return to Fct. 3.08 “COM”.

4/5