KROHNE TIDALFLUX IFM 4110 PF EN User Manual

Page 51

Advertising
background image

7.30830.32.00

51

7.4

Hardware test, Fct. 2.03

Please note:
Before beginning the test, deactivate any alarms and controllers as the current output will be tested with test values 4, 4.7 and
23 mA for a short period.
Key

Display

Description

if “YES” is selected in Fct. 3.04 ENTRY CODE,
key in 9-stroke CODE 1 now:

.

Fct. 1.00

OPERATION

Fct. 2.00

TEST

Fct. 2.01

TEST Q

2x

Fct. 2.03

HARDW. TEST

Hardware test

SURE NO

SURE YES

WAIT

Hardware test in progress, duration approx. 60 seconds

- - - - - - - -

1st error

List of errors see sect. 4.5. Errors are

- - - - - - - -

2nd error

always displayed independent of setting in

- - - - - - - -

3rd error

Fct. 1.04. If no error is detected, refer to next line.

Fct. 2.03

HARDW. TEST

Terminate hardware test

(2x) 3x

- - - - - - - -

- - - - - / - - -

back to measuring mode

If you need to return your flowmeter to Krohne, please refer to the last-but-one page of these instructions

7.5

Faults and symptoms during start-up and flow measurement

Most faults and symptoms occurring with the flowmeters can be eliminated by following the instructions indicated in the

following tables.

For greater clarity, faults and symptoms in the tables are divided into different groups:

- LED light-emitting diodes on the front panel (status messages)
- D

display

- I

current output I

- P

pulse outputs P and A1

- S

status outputs D1, D2, A1 and A2

- C

control inputs C1 and C2

Before contacting the Krohne Service Department, please read the instructions in the table.

Group LED

Display

Cause

Remedial action

A/D converter range exceeded

Check if setting in Fct.
3.06 (menu “FLOW”) is
“PULSATING”.

Reduce flow rate; if
unsuccessful, test as
described in sect. 7.6

Level too low in measuring tube

Fill measuring tube up to at least
10 percent, see also sect. 7.3.

LED 1

Both LED’s flash

Problems in primary head

No communication between
primary head IFS 4000 PF
and signal converter IFC
110 PF, check data cable
according to sect. 1.5.6.

General failure in primary
head, see sect. 7.6