Modbus protocol, 7 zero calibration – KROHNE UFC 400 Modbus EN User Manual
Page 19

MODBUS PROTOCOL
5
19
UFC 400
www.krohne.com
05/2013 - 4002744801 - AD Modbus UFC 400 R01 en
5.8.7 Zero Calibration
A zero calibration is not necessary for normal operation. It's recommended that only trained
service engineers should perform a zero calibration.
The Coils and Registers below are required to perform a zero calibration. For further
information refer to
Zero Calibration
on page 24.
Modbus Function Codes "Read Coils" (0x01) and "Write Single Coil" (0x05)
Coil
Address
Name
Function
Action
Values
Display
Fct. No.
2000
2000
2000
2000
(0x07D0)
Zero Calibration
Control
Start
Write
1 = start zero calibration
C1.2.1
Status
Read
0 = zero calibration not running /
complete
1 = zero calibration in progress
-
Modbus Function Code "Read Input Registers" (0x04)
Input
Register
Name
Type
No. of
registers
Values
20000
20000
20000
20000
(0x4E20)
Zero Calibration Result
Byte
1
1 = OK
0 = NOK