Modbus protocol, 9 modbus rs485 communication settings – KROHNE UFC 400 Modbus EN User Manual
Page 21

MODBUS PROTOCOL
5
21
UFC 400
www.krohne.com
05/2013 - 4002744801 - AD Modbus UFC 400 R01 en
5.8.9 Modbus RS485 Communication Settings
Modbus Function Codes "Read Holding Registers" (0x03) and "Write Multiple Registers" (0x10)
Holding
Register
Name
Description
Type
No. of
registers
Values
Display
Fct. No.
50000
50000
50000
50000
(0xC350)
Baud rate
Baud rate for the RS485
Modbus Communication in
[bps]
Long
2
1200, 2400, 3600,
4800, 9600, 19200
(default), 38400,
57600, 115200
C5.8.2
50002
50002
50002
50002
(0xC352)
Transmission
Delay
Select transmission delay
between Modbus request and
Modbus response in [ms]
Float
2
0 (default)...0.04
C5.8.5
50004
50004
50004
50004
(0xC354)
Slave Address
Modbus Slave Address
Byte
1
1 (default)...247
C5.8.1
50005
50005
50005
50005
(0xC355)
Parity
Parity for the RS485 Modbus
Communication
Byte
1
0 = Even (default)
1 = Odd
3 = Off
C5.8.3
50006
50006
50006
50006
(0xC356)
Register
Format
Switch Register Format Little
Endian or Big Endian
Byte
1
0 = Little Endian
1 = Big Endian
(default)
C5.8.4
50007
50007
50007
50007
(0xC357)
Stop Bits
Number of Stop Bits
Byte
1
1 = 1 Stop Bit
(default)
2 = 2 Stop Bits
C5.8.6