KROHNE GFM 700 EN User Manual

Page 17

Advertising
background image

17


Fct. Text

Description

and settings

3.5.0 USER DATA

Submenu 3.5.0 User data

3.5.1 LANGUAGE Language

for display texts, see

Sect. 5.11

GB/USA (English)

D (german)

F (French)

N (Dutch)

3.5.2 ENTRY.CODE.

1

Entry code 1 for setting level

required?

See Sect. 5.12

NO = Entry with

 key only

YES = Entry with

 key and

9-keystroke code

Setting of the code under Fct. 3.5.3

3.5.3 CODE 1

Set Code 1, see Sect. 5.12

(appears only if YES set under Fct.

3.6.2)

• Factory setting:

→, →, →, ↵, ↵. ↵, ↑, ↑, ↑

• Different code required:

Press any 9-keystroke combination

and then press the same

combination again.

Each keystroke acknowledged by

*

”.

WRONG CODE appears if 1

st

and

2

nd

entries are not equal.

Press

↵ + → keys

and repeat entries.

3.5.4 LOCATION

Tag name setting

(measurement point no.)

max. 10 digits. Required only

for flowmeters of “HHC” design

(operator control via Hand-Held

Communicator MIC 500, connected

to current output).

Factory setting: ALTOMETER

Characters assignable to each

place: A..Z / a..z / 0..9 / + / -./

underscore character =

blank character.

3.5.5 OUTP. HOLD

Hold values of outputs

during settings?

Setting: NO or YES

3.5.6 UNIT

TEXT

Text

for user-defined unit

Factory setting:

hLiter/hr or US.MGal/DAY

Characters assignable

to each place:

A..Z / a..z / 0..9 / + / −-./

underscore

character = blank character.

Fraction bar ”/” in 7

th

place is

unalternable.

3.5.7 FACT. QUANT Conversion factor for quantity F

M

F

M

= quantity per 1 m

3

!

Factory setting:

1.00000 E1 (for hecto Liter)

or 2.64172 E-4 (for US M.gallons)

Value setting:

0.00001*10

9

- 9.99999*10

+9

3.5.8 FACT. TIME

Conversion factor for time F

T

F

T

in seconds!

Factory setting:

3.60000 E3 (for hour) or

8.64000 E4 (for day)

Value setting:

0.00001*10

9

- 9.99999*10

+9


Advertising