7.2 Error

messages

Fct. Text

Description

and settings

4.0

PARAM.

ERROR

Main menu 4.0 Parameter error

4.1.0

FLOW

VELOC.

FLOW VELOCITY “v” incorrect:

Ensure condition

0.5 m/s

v 60 m/s

or 1.5 ft/s

v 180 ft/s is met!

4.1.1

FULL SCALE Full-scale range for flowrate Q

100%

,

see Fct. 3.1.1

4.1.2

METER SIZE Meter size, see Fct. 3.1.5

4.2.0

F/R FLOW

FULL-SCALE RANGE(S) for

forward/reverse flow incorrect:

Ensure condition F

R is met!

4.2.1

FULL SCALE Full-scale range for forward

flowrate Q

F100%

, see Fct. 3.1.1

4.2.2

REV. SCALE Different range for reverse flow

required? See Fct. 3.1.2

4.2.3 REV.

VALUE

Full-scale range for reverse flow

Q

R100%

, see Fct. 3.1.3

4.3.0

I RANGE

CURRENT OUTPUT I RANGE

incorrect:

Ensure condition

I

100%

I

0%

4 mA is met!

4.3.1

I 0 PCT

Current for 0% flow (I

0%

), see Fct.

3.3.3

4.3.2

I 100 PCT

Current for 100% flow (I

100%

), see

Fct. 3.3.4

4.4.0

I MAXIMUM

CURRENT LIMITATION

incorrect:

Ensure condition

I

max

I

100%

is met!

4.4.1

I 100 PCT

Current for 100% flow (I

100%

), see

Fct. 3.3.4

4.4.2

I MAX mA

Setting of max. output current

(I

max

),

see Fct. 3.3.5

4.5.0

LFC. I RANG. LOW-FLOW CUTOFF RANGE I

incorrect:

Ensure condition

cutoff “off”.

Cutoff “on”

1% is met!

4.5.1 L.F.

CUTOFF

I

Low-flow cutoff (SMU) for current

output required? See Fct. 3.3.7

4.5.2

CUTOFF ON Cutoff “on” value SMU-I, see Fct.

3.3.8

4.5.3 CUTOFF

OFF

Cutoff “off” value SMU-I, see Fct.

3.3.9

4.6.0

LFC. F

RANG.

LOW-FLOW CUTOFF RANGE F

incorrect:

Ensure condition cutoff “off”.

Cutoff “on”

1% is met!

4.6.1 L.F.

CUTOFF

F

Low-flow cutoff (SMU) for

frequency output required?

See Fct. 3.4.6

4.6.2

CUTOFF ON Cutoff “on” value SMU-F,

see Fct. 3.4.7

4.6.3 CUTOFF

OFF

Cutoff “off” value SMU-F,

see Fct. 3.4.8