4 environmental conditions – KROHNE GFM 700 EN User Manual

Page 9

Advertising
background image

9

Local display

3-field LCD

Display function

actual flowrate, measurement of sound velocity to determine (compo-
sition of) liquid product, forward, reverse and sum totalizer (7-digit) and
status messages; each can be set for continuous or sequential display

Units: actual

flowrate

m

3

, liter, US gallons per second, minute or hour,

or in user-defined unit, e.g. liter/h or US gallon/day

totalizer

m

3

, liter, US gallons or in user-defined unit, e.g. hecto liter or US million

gallon (min. 1 year overflow time)

Language of plain texts

English, French, German, Dutch, other on request

Display 1st line (top)

8-character, 7-segment numeral and sign display,
and symbols for key acknowledgement

2nd line (middle)

10-character, 14-segment text display

3rd line (bottom)

5 markers to identify display in measuring mode

Housing
Material

die-cast aluminium with polyurethane finish

Protection category

IP 67 (IEC 529 / EN 60529) equivalent to NEMA 4 and 4X

Signal cable

only for remote systems (F)

length up to 10 m / 30 ft (max. 20 m / 60 ft, option)


4.4 Environmental conditions

In accordance with EN IEC 61010-1 the following enviromental conditions have to be observed .


The GFM 700 is designed to operate safe under the following conditions:
a)

suitable for indoor and outdoor use, the housing of the signal converter is usable up to Protection Category IP67

b)

IP 67 is only warranted when using suitable cabling in the cable glands

c)

use up to an altitude of 2000 m above see level

d)

suitable for an operation ambient temperature range - 40 …. +60 °C

e)

suitable for an storage temperature range -40 …. + 80 °C

f)

suitable for use in atmospheres with a relative humidity up op 80%

g)

mains supply voltage fluctuations up to +

13 % of the nominal voltage range

h)

withstands over voltages up to category II on the main supply voltage ( IEC 60364-4-443)

i)

connected to protective earth conductor ( Protection Class I)

j)

rated pollution degree 2

Advertising