Operation, 3 operation mode – KROHNE OPTIWAVE 5200C SIL User Manual
Page 14

7
OPERATION
14
OPTIWAVE 5200 C
www.krohne.com
02/2013 - 4001906202 - AD SIL OPTIWAVE 5200 R02 en
7.3 Operation mode
In operational conditions, the device continuously measures the distance between the tank
connecting flange face / thread stop and the surface of the product.
The operation mode of the device is shown in the flow chart that follows:
1
Initial state (start the device / new measurement)
2
Make sure that sufficient energy is available
3
Is there sufficient energy?
3a
No. There is not sufficient energy.
3b
Yes. There is sufficient energy.
4
The device does an auto-diagnostic check at
intervals of 24 hours
5
Is there an error condition?
5a
Yes. There is an error conditon.
5b
No. There is not an error condition.
6
Distance measurement
7
Is a correct measurement available?
7a
No. A correct meaurement is not available.
7b
Yes. A correct meaurement is available.
8
Error mode state
9
Current output (distance value or a derived
measurement value)
10
Is the current output setting correct?
10a
No. The current output setting is incorrect.
10b
Yes. The current output setting is correct.