Service, 5 calibration procedure – KROHNE OPTIWAVE 5200C SIL User Manual
Page 23
SERVICE
8
23
OPTIWAVE 5200 C
www.krohne.com
02/2013 - 4001906202 - AD SIL OPTIWAVE 5200 R02 en
8.5 Calibration procedure
8.5.1 General notes
If the antenna is not in the same customer order as the signal converter, it is necessary to
calibrate the device.
Equipment needed
• A device with the integrated display option.
• Process measurement and device configuration software (e.g. the DTM for PACTware™), if
the device does not have the integrated display option.
• An ammeter.
• Reference device: an approved level meter or indicator
Requirements
• The ammeter must be correctly calibrated and connected directly to device.
• You must have 2 reference points (levels) in the tank given by a different measurement
solution. These points are identified as "ref_pt1" and ref_pt2".
CAUTION!
Do not connect the ammeter across the safety system PLC because this configuration can open
the circuit breaker.
Make sure that the 2 reference points are not in the top or bottom dead zones.