4-20 mA Configurations

Device Settings: (Refer to Table 1 for details of this output option.)
1) Using a small flat blade screwdriver, set the 10-position selector

switch to position 0.

2) Verify that the 2-position jumper JP1 is shunting pins 2 & 3.
3) Connect the output connector to position B, as shown in Figure 3.

Table 3: (Field configurable settings for 4-20mA output option)
Configurable Control:

Setting:

10 position selector switch

Position 0

2 position jumper JP1

Pins 2 & 3 shunted

Output connector

Position B

Input Voltage:
The supply voltage must be between 12 and 35 VDC. The maximum

resistance that may be placed within the current loop is given by the

following formula:

Rmax = 50(Vs - 12)

Rmax = the maximum resistance that may be placed in the current loop (?)

Vs = the value of the supply volt age (VDC)

NOTE: Although the signal

conditioning circuit does have

integral over-current protection,

manufacturer suggests that the

circuit be protected with a 0.25

amp fuse.
1) Connect the positive DC

power source (+12 to +35

VDC) to terminal # 1 on the

DIN connector or the red

wire on the pigtail.

2) Connect terminal #2 of the DIN connecter or the black wire from the

pigtail to the positive current input on the receiving device.

3) If the power source does not originate from the receiving device, the

negative side of the power supply must be connected to the signal

ground of the receiving device.

4) If the transmitter is operating properly, the green LED on the signal

conditioning board will illuminate dimly at zero flow and will increase

in intensity as flow increases.

Configuration / Operating Modes

Field Configurable Outputs
The FreeFlow meter can be field configured to of fer a variety of output

protocol signals. Possible outputs protocol options are 4-20mA, 0-5Vdc

and 0-10Vdc. The following steps must be taken prior to making any

field configuration changes.
1) Fluid flow to the device should be shut of f; the device must not be

under pressure.

2) The device should be disconnected from power prior to any field

calibration.

3) All safety protocols for the environment in which the device is

installed must be considered; remove the device from the inst allation

if needed.

4) Removing the cover will reveal the internal signal conditioning

circuitry and the field configurable controls. Figure 1 illustrates the

field configurable attributes of the device which shall be referenced

in subsequent sections of this manual.