NOSHOK 5301 Series Load Pins User Manual

Page 3

Advertising
background image

Operating manual

5301 Series Load Pins

3

Maintenance

The force transducers in the F5301 model series are maintenance-

free.

During welding work the transducer must be bypassed with a copper

wire (min. 50 mm

2

) so that welding current does not flow through the

transducer and weld the force introduction points.

NOSHOK, Inc.

1010 West Bagley Road, Berea, OH 44017

Ph: 440.243.0888

Fax 440.243.3472

www.noshok.com

Ambient temperature

The temperature range of -20°C to +80°C that is specified in the data

sheet applies with regard to deployment. The specified error limits are

not guaranteed outside this temperature range.

Moisture and corrosion protection

Tropical climates and condensation are not a problem because the

transducers comply with protection class IP 67 in accordance with

EN 60529:1991+A1:2000 / IEC 529.

Deposits

Dust, dirt and other object must not be allowed to form deposits in

such a way that they create a force short-circuit to the measuring

spring, which would falsify the measuring signal.

Precautions to take during assembly

• Force measuring equipment is extremely sensitive and must be

handled carefully.

• Attention must be paid to the installation position and therefore the

load direction when the force transducers are being installed.

• The transverse loads and side forces also include the relevant

components of the measured variables, which may be introduced at

an incline.

• Care must therefore be taken to keep the transducer free of lateral

loads and torsion during assembly (e.g. when tightening the lock

nuts).

• Overloading must be prevented at all times.

• The output signal (c=1-2mV/V, 4...20mA, 0…20mA, 0...10V) is noted

on the name plate (Fig. 1). The connection assignments are also

noted on the name plate and can also be found in the section entitled

“Electrical connection”. The polarity must be correct at all times.

SEE NEXT PAGE FOR NAME PLATE DIAGRAM.