Please note the following:

• Always use shielded, low-capacity measuring cables

(all tecsis cables meet these requirements, see chapter 10.1).

• Do not route the measuring cable parallel to high-voltage current and

control cables.

• Avoid leakage fields from transformers, motors and contactors.

• The transducer, the amplifier and the display unit must not have

multiple grounds. Attach all equipment to the same protective

conductor.

SIL-3 Elektronic

Amplifier-Electronics 4...20mA or 0...10V

for SIL-3 applications with 2-channel PC control

NOSHOK, Inc.

1010 West Bagley Road, Berea, OH 44017

Ph: 440.243.0888

Fax 440.243.3472

www.noshok.com

This well proven design has been amended by an additional resistor R7

in order to monitor the condition of the amplifier unit and signal path.

This resistor is connected as a shunt to resistor R5 by a relay contact

(a) as soon as an excitation voltage U

r

appears at relay A.

a

R5

R6

R4

R3

R2

R1

A

Ur

R7

Ub

U0

The connection of resistor R7 will always result in a defined unbalancing

of the zero point (diagonal voltage) of the Wheatstone Bridge.

An external independent control unit activates relay A which changes

the output by a certain value. Because of security reasons the control

unit has to be a 2-channel one. When the expected change of the

output signal is detected it can be assumed that the whole signal path

(Wheatstone Bridge – amplifier – output) works well.

If it does not appear it can be concluded that there is a defect in the

signal path.

The standard adjustment of force transducers with current output for

overload control is e.g.: