101 MAGNETIC FLOW METER INSTRUCTIONS

www.seametrics.com

Mounting. Mount the control housing to a secure

surface with screws or bolts. To gain access to the

mounting holes, remove the front cover. The mounting

holes are directly under the front cover screws. The

primary element is supplied with a foot bracket, which

can be attached to a secure surface. Alternatively, the

unit can be supported by the piping and the foot bracket

can be removed.

Connections. The unit ships with the coil activation and

signal leads already connected to the control housing. To

connect output signal or alarm relay leads, remove the

front cover. See the “Connections” diagram. Power

connection uses a standard power cord. If conduit

connection is required, remove the cord and strain relief

and use the strain relief hole for a conduit connector.

Grounding. Important: For proper operation, one or both

of the ground lugs must be well connected to a good

INSTALLATION

CAUTION:

Although this meter has an

empty pipe detection function, under
certain conditions of empty or
partially-full pipe the meter may read
a flow when there is none. If this is a

hazardous condition, mount the meter in such a way
as to ensure the meter will always be full of liquid.

quality earth ground. (The ground lugs also retain the

foot bracket). See the Grounding diagram on page 4.

All magmeters come with a 12 feet ground wire attached.

OPERATION

Display. The flow rate is displayed in the time and

volume units which have been selected (liters/min, for

example). The cumulative total flow is displayed in the

chosen units, up to eight digits. It then resets to zero

and begins again.

Outputs. The analog output which has been chosen

varies continuously with the flow. If the output is too

“jumpy” (changes too frequently), it can be damped

either by increasing the averaging time (see the “Fast

Analog Output” setting) or by selecting “Disabled”

under Fast Analog Output and increasing the amount

of damping using the Low Pass Filter setting. The pulse

output will produce a 50% duty cycle pulse at the

volume intervals for which it is set - one pulse per

liter, for example. Note that since each pulse consists

of equal times on and off, if the interval between

pulses is large the pulse may remain in the “on”

condition for several seconds. The relay alarm output

will only energize if the flow goes above (high) or

below (low) the flow alarm setting. The alarm relay

will remain energized until the flow exceeds the

setpoint by .25% (hysteresis).