EM SERIES

101 MAGNETIC FLOW METER INSTRUCTIONS

1- 800 -975 - 8153

SeaMetrics Inc.
EM101

WAIT 2 SECONDS

Rate
Tot

Volume Units G

Time Units M

Tot Dec Point
Format 0.0

An. output Type

Fast An. Output

Rate For Analog
Output:

Rate For High
Alarm:

Rate For Low
Alarm:

Pulse Output

Low Pass
Digital Filter:

Totalizer Reset

SeaMetrics Inc.
EM101

SETTING

D I S P L A Y S

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

This is the power-up display.

Rate/total display (shows two seconds after power-up), Press to begin programming

Use the key to choose the volume unit: milliliters (mL) liters (L) or gallons (G)

Use the key to choose the time base for rate: per second (S) per minute (M) per hour

(H) per day (D)

Use the key to choose number of decimal places: 0, 0.0, 0.00

Leave “Analog Output Type” unchanged unless you intend to use the 0-5 V output.

Standard setting is for 4-20 mA. Consult factory for 0-5 V output.

Leave “Fast Analog Output” unchanged unless you have a demanding closed-loop application.

Standard setting is “disabled”. If you enable this output, use the key to select a response

time, 50 to 1000 millisecond. This will control how rapidly the analog output tracks flow.

Use the and key to set the percentage of full scale at which peak analog output

(20 mA or 5 V) occurs. On a max 10 gpm meter, 60% would be 6 gpm, for instance.

Use the and key to set the percent of full scale at which the high flow alarm will

energize. On a 10 gpm meter, 90% would equal 9 gpm, for instance.

Use the and key to set the percent of full scale at which the low flow alarm will

energize. On a 10 gpm meter, 10% would equal 1 gpm, for instance.

Use the and key to set the number of pulses (0000 to 9999) per unit. If you have

selected liters, it will be pulses per liter, etc.

This setting controls the smooth/responsive trade off of the display. Increasing the time makes

the display smoother, but slower to respond to change. Use the to set. If using the meter

with pulsing flows, it may be necessary to increase the time. Choice: 1, 2, 5, 10, 20 seconds.

Use the key if it you wish to enable the totalizer reset. It will switch between “Enabled”

and “Disabled”. If the reset is enabled, every time the key is pressed (during normal

operation) the totalizer will reset to zero.

Settings finished. Wait five seconds to return to normal operations