The following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments directly for further assistance.

The DRI does not turn on:

• Check all electrical components (power source … battery or power cord)

The DRI does not recognize a sensor:

• Check connection between the DRI and the remote sensor

• Refer to calibration instructions for a sensor that cannot be recognized

Error Codes are displayed:

• If any error codes appear on the display, turn power OFF and then ON

• If error code remains, see table below:

Troubleshooting

Calibration (continued)

DISPLAY

CONDITION

ACTION

EE PROM reading error

Press RESET; unit will default to
factory settings

Memory data error

Press RESET to clear memory

Calibration error

Press RESET and repeat calibration

procedure

Memory unit error; occurs if the

Change measurement units or clear

measurement unit (lb, Kg, N) is

memory

changed while entering data into

memory
Sensor error; occurs if the remote Change sensor or clear memory
remote sensor is changed during

data entry to memory
Overload condition –possible

Remove excessive load; if the display

load cell damage

does not return to normal operation,

send unit in for repair

MODEL

SELECT NUMBER

DISPLAY CALIBRATION WEIGHT

LC-50S

06

50L

25kg

LC-100S

07

100L

50kg

LC-200S

08

200L

100kg

LC-500S

10

500L

250kg

TS-3SD, TS-3J

18

t50o

TS-10SD, TS-10J

19

t10L

TS-20SD, TS-20J

20

t20L

TS-50SD, TS-50J

21

t50L

TS-100SD, TS-100J 22

t100L

Please call Shimpo for information on additional sensors.