Operation (continued) – SHIMPO DRI-series User Manual

Page 8

Advertising
background image

8

Standard Memory Mode
After selecting the “nonE” mode it is possible to store the maximum “+”, maximum “-”, minimum “+”, minimum “-”, and last reading
during a sampling period. After pressing the MEMO SET button, a flashing “M” will appear on the small display; push the MEMO SET
button again to end the sampling period.

Each time the RECALL button is pressed the DRI will scroll through the readings mentioned above. It is not necessary to clear the
memory before performing the next test as they will be cancelled when the MEMO SET button is pressed.

Recalling Statistics

While recalling data from memory in either the Single or Continuous Mode it is possible to also recall statistics. To recall statistics:

1. Press the RECALL button. The last data reading entered and memory position will be flashing opposite each other on the main

display.

2. Press the PEAK button. The maximum reading will appear on the main display.

3. Each subsequent push of the Peak button will scroll to minimum, peak “+”, peak “-”, average, standard deviation and return to

maximum.

4. Press the RESET button to exit.

NOTE: Peak “+” and peak “-” will not be displayed if the data was stored in the Single Memory Mode.

Setting of Filter

Subtle vibrations are common in many testing environments. The DRI incorporates a filtering feature that “smoothes” the force curve by
eliminating extraneous measurements caused by these outside influences. Depending on how severely these impact the sampling, a large
or small filter is available. Set a filter as follows:

1. Press the PEAK button while holding down the ON button. The small display will show “dSP”.

2. Press the PEAK button eight times. The small display will indicate “Adf” and the main display will reflect the current setting.

3. Each time the > button is pressed, the main display will scroll through “0” (no filter), “1” (small filter) and “2” (large filter).

4. Press the MEMO SET button to store the selection and exit.

NOTE: To determine which filter (if necessary) is needed, first attempt to test without a filter, then proceed to filter 1 and possibly to filter 2.

Operation (continued)

Advertising