INSTRUMENT SETTINGS


Temperature Log Recording
To record the temperature logs (block temperature
profile) of each run, touch the button to set this to
Always. If temperature logs are not required, set to

Never.

Auto Resume On Power Loss
This determines the time limit after which the unit
will restart after a power failure. For example, if the

Auto Resume is set to 30 minutes, the unit will
restart at the beginning of the step at which it was
interrupted if the power failure lasted for 30 minutes
or less but not if it was for longer.

Setting Auto Resume to Always will always allow

the unit to re-start irrespective of the length of the
power failure.

Date and Time Settings
Use these settings to set the current date, time and

date format on the unit.

• When set, touch Back to return to the Settings

menu.







Date

• Touch Set Date in order to enter the current

date.

• Use the + and keys to set the date values.


• When the correct date has been set, touch OK

to return to the previous screen.

• Touch Cancel to return to the previous screen if

no changes are required.