Techne Prime Elite User Manual

Page 49

Advertising
background image

47

Note: If the user attempts to send a program to unit
while it is already running, the warning screen

shown opposite will be displayed.

• Touch OK to exit.








THE PRE-RUN SCREEN


The Pre-run Screen shows a graphical representation of the program allowing for a rapid check of the
temperature parameters.

The centre of the screen shows each temperature
step and hold time (not to scale). Stages are

indicated by a white bar spanning the steps within
the stage.

The bottom of the screen shows the program name,
the date and the expected end time if it was started

immediately.

• Touch Start to start the program.

• Touch Edit to edit the program.

• Touch Back to return to the previous screen.

Note: The temperature log will remain blank until

the program has been started.

If a step has a gradient programmed into it, the bar
will be shown thicker and indicate the upper and
lower temperatures of the gradient range.

If a step has an incremented or decremented
temperature, this will be indicated by a white spot
next to the temperature.

If a step has an incremented or decremented time,

this will be indicated by a white spot next to the
time.








Gradient

Inc/dec

temperature

Inc/dec time

Gradient

Inc/dec

temperature

Inc/dec time

Advertising