Instruction Manual CON 6/TDS 6

25

6

ADVANCED SETUP FUNCTIONS

6.1

Advanced Setup Overview

The advanced setup mode lets you customize your meter‟s preferences and

defaults. To enter advanced setup mode:
1)

Make sure that the meter is switched-off.

2)

Press ON and MODE key simultaneously, holding both keys for 2

seconds. First release ON key first before releasing the MODE key.

3)

[StUP] indicator will appear momentarily and [CELC] will appear next.

4)

Overviews of CON 6 and TDS 6‟s Setup Menu as follows.


Enter Setup Page.

Select Cell Constant. Choice of k = 0.1, 1.0,

and 10.0.
Default value is 1.0.

Select Automatic Calibration. “Yes” for auto

calibration and “no” for manual calibration.
Default value is “Yes”. (Available in CON 6
meter only)

%

Adjust Temperature Coefficient value from 0.0

to 3.0 %/°C.
Default value is 2.1 %/°C.