ADVANCED SETUP FUNCTIONS
6.1
Advanced Setup Overview
The advanced setup mode lets you customize your meter‟s preferences and
defaults. To enter advanced setup mode:
1)
Make sure that the meter is switched-off.
2)
Press ON and MODE key simultaneously, holding both keys for 2
seconds. First release ON key first before releasing the MODE key.
3)
[StUP] indicator will appear momentarily and [CELC] will appear next.
4)
Overviews of CON 6 and TDS 6‟s Setup Menu as follows.
Enter Setup Page.
Select Cell Constant. Choice of k = 0.1, 1.0,
and 10.0.
Default value is 1.0.
Select Automatic Calibration. “Yes” for auto
calibration and “no” for manual calibration.
Default value is “Yes”. (Available in CON 6
meter only)
%
Adjust Temperature Coefficient value from 0.0
to 3.0 %/°C.
Default value is 2.1 %/°C.