14 addendum 2: calculating tds conversion factor – Thermo Fisher Scientific EcoScan CON 6 & TDS 6 User Manual

Page 50

Advertising
background image

Instruction Manual CON 6/TDS 6

46

14 ADDENDUM 2: CALCULATING TDS CONVERSION

FACTOR

You can calibrate your meter using TDS calibration standard solutions. The

calibration standard only needs to give the TDS value at a standard

temperature such as 25 °C. To determine the conductivity-to-TDS conversion

factor use the following formula:

Factor = Actual TDS ÷ Actual Conductivity @ 25 °C

Definitions:

Actual TDS: Value from the solution bottle label or as a
standard you make using high purity water and precisely
weighed salts.

Actual Conductivity: Value measured using a properly
calibrated Conductivity/Temperature meter.


Both the Actual TDS and the Actual Conductivity values must be in the same

magnitude of units. For example, if the TDS value is in ppm the conductivity

value must be in µS; if the TDS value is in ppt the conductivity value must be in

mS.

Check your factor by multiplying the conductivity reading by the factor in the

above formula. The result should be in TDS value.

Advertising