Instruction Manual CON 6/TDS 6

36

9

ERROR MESSAGES

LCD Display

Indicates

Cause

Solution

“LO” indicator

appears.

Low battery level.

Need new batteries

or battery

connection is bad.

Clean battery

contacts. Replace

batteries with fresh

ones, noting polarity.

Err 1

Conductivity

calibration error

Calibration point is

outside the ±40%

window in the auto-

calibration.

Check the value of the

conductivity calibration

solution.
Switch to manual

calibration mode and

calibrate again.
If message persists,

return unit*.

Err. 2

Temperature

calibration error.

Auto calibration is

performed outside

the temperature

range (0 – 50 °C).

Check the temperature

and make sure that it

is within the

acceptable range.
If message persists,

return unit*.

Err. 3

Conductivity

calibration error.

Calibration point is

within 10% of the

measurement

range in the

manual calibration

mode.

Check the value of the

conductivity calibration

solution.
If message persists,

return unit*.


* See Sections on “Warranty” and “Return of Items”.
If an error message appears, switching off the meter and switching it on again

may eliminate the error message. Refer to diagram on right.
If error persists, or the meter shows incorrect values,

return the meter.
For a complete diagram of the display see page 3.