7

PROBE CARE AND MAINTENANCE

Keep the conductivity probe clean. Rinse the probe twice, and gently swirl it

while you take readings. For best accuracy, soak a dry probe for at least 5 to 10

minutes or longer before calibration. Rinse the probe with deionized or tap

water before storing. Never scratch the bands with a hard substance. Do not

strike the probe against any hard surface.

Do not immerse the probe in oily solutions. Clean the

electrode thoroughly by stirring it in a mild detergent

bath or isopropyl alcohol. Wipe the probe with a soft

tissue paper. Rinse thoroughly in tap water and then in

deionized water. Recalibrate the meter after cleaning

the probe.

The conductivity probe (Order Part No. EC-

CONSEN91B/ 35606-55) which is included with your

meter features a removable probe guard to make

cleaning easy.

To remove probe guard:
1)

Grip yellow probe guard and twist clockwise. The

locking notch will release.

2)

Slide probe guard off end of probe.


NOTE: Remember to re-attach the probe guard prior to taking readings. Failure

to do so could result in erroneous readings.