Sp1 sp2, S.cd – Thermo Fisher Scientific Alpha pH 1000 User Manual

Page 12

Advertising
background image

7

SETUP

HOLD

000

S.Cd

4.2 Function

Groups

The main function and sub-function groups are organised in a matrix format for configuration and
selection of parameters. The main function groups are:


1) Offset

adjustment

(OFS)

2)

Temperature Measurement / compensation settings (Set

o

C)

3)

Control relay 1 configuration (SP1)

4)

Control relay 2 configuration (SP2)

5)

Control type (Cntr)

6)

Current output (mg)

7) Configuration

(ConF)

8)

Calibration (CAL pH)

HOLD

SETUP

HOLD

SETUP

HOLD

HOLD

SETUP

SETUP

Sp1

Sp2

HOLD

SETUP

°

C

HOLD

SETUP

Cal

HOLD

SETUP

HOLD

SETUP

C

OFS

The set-up parameters can be viewed or changed by entering a security code. See section 5.2
for security code information.

4.2.1

How to view operating parameters without access to change them:

a) Press

the

Enter key. The display will prompt the user to

enter a security code (S.Cd). Leave the security code at
“000” (do not enter a security code).

b)

Press ENTER key again. This allows you to only view (not
change) any sub-functions’ settings.

c)

Press the

▲ or ▼ key to scroll through the sub-functions.

d)

Press the ENTER key at a particular sub-function to view in detail.

e)

Press the ENTER key to return to the sub-function menu.

f)

Press the

▲ or ▼ keys simultaneously (as an Escape key) at any time to return to the

Measurement mode.

Note: To simplify operations, the controller will not display parameters that are not relevant to a
particular subfunction. For example, if the user set the controller for limit control, it will not display
pulse length/frequency settings.