23

7.6.6

Configuring the Wash contact if relay 3 has been set to the Wash mode

1) Follow direction in 7.6.1 to enter Configuration mode.
2) Press the ENTER key until upper display shows “Alr” or

“CLn”.

3) Press the

or key to select the “CLn” function. Press the

ENTER key.

4) Press

the

or key to select the wash cycle (int. 0.1 to

199.9 hours) and press ENTER.

5) Press the

or key to select the wash duration (1 to 1999

seconds) and press ENTER.

Note: During wash cycle, the controller is set to HOLD. For safety
reasons, the HOLD function starts 5 seconds before the start of
wash cycle and will continue for 10 seconds after the wash cycle is
completed.




7.6.7

Selecting pH sensor type

Note: This parameter is blocked out in ORP mode. Proceed to 7.6.6.

Note: After each changeover, the factory calibration data for zero point and slope are
overwritten. Be sure to recalibrate (see section 6).

This parameter group lets you configure the controller for glass or antimony electrodes. Use an
antimony electrode for systems with hydrofluoric acid content.

1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the

▲ or ▼ key until the

upper display shows “GLAS” or “AnTY” and lower display
shows “EL”.

-

EL = electrode

- GLAS

=

glass

- AnTY

=

antimony

3) Press the

or key to select glass or antimony.

4) Press the ENTER key to confirm your selection.
5)

Proceed to 7.6.6 step 3, or return to Measurement mode by

pressing the

▲ and ▼ keys simultaneously (escape).

SETUP

HOLD

10.0

SETUP

HOLD

10

SETUP

HOLD

Glas

SETUP

HOLD

El

SETUP

HOLD

El