The output relay of the pulse length controller is clock-timed. The switching period T remains
constant. Depending on the divergence from the limit value, the switch-on time t

ON

is increased

or decreased in accordance with the proportional range Xp.

The following applies:

t

ON

+ t

OFF

= T (Const.)

greater divergence

→ greater t

ON

Xp

exceeded

→ t

ON

= T (relay remains picked up)

Relay

ON

tON

tOFF

Pulse Length T

OFF

Time [s]


Controller signal of Pulse Frequency Controllers
The output relay of the pulse frequency controller is clock-timed. The pulse duration t

ON

remains

constant at 250mS.

Depending on the divergence from the limit value, the frequency (

1

/

T

) is increased or decreased

in accordance with the proportional range Xp.

The following applies:

t

ON

= Const. (250 msec.)

greater divergence

→ greater f (greater frequency)

Xp exceeded

→ max. frequency