9 technical specifications – Thermo Fisher Scientific Alpha pH 1000 User Manual

Page 31

Advertising
background image

26

9 TECHNICAL

SPECIFICATIONS

pH Range

0.00 to 14.00 pH

Resolution

0.01 pH

Relative Accuracy

± 0.01 pH

mV Range

0 to 100.0% / -1000 to +1000 mV

Resolution

0.1% / 1mV

Relative Accuracy

± 1 mV

Temperature

- 9.9 to + 125.0

o

C

Resolution

0.1

o

C

Relative Accuracy

± 0.5

o

C

Sensor

Pt 100 /Pt 1000 (jumper selectable)

Temperature Compensation

Auto / manual (reference at 25.0

o

C)

Set-point and Controller Functions

limit controller

Function (switchable)

pulse length/ frequency controller

Controller characteristics

proporational controller

Adjustable period with pulse length controller

0.5 to 20 sec.

Adjustable period with pulse frequency controller

60 to 120 pulses/min

Pickup / Dropout delay

0 to 2000 sec.

Switching pH hysteresis

0.1 to 1 pH

Switching ORP hysteresis

1 to 10.0% / 10 to 100 mV

Contact outputs, controller

2 potential-free change-over contacts

Switching voltage / current / power

max. 250 VAC / max. 3A / max. 600 VA

Alarm Functions

Function (switchable)

Latching / pulse

Wash cycle

0.1 to 199.9 hrs

Wash duration

1 to 1999 s

Pickup delay

0 to 2000 s

Switching voltage / current / power

max. 250 VAC / max. 3A / max. 600 VA

Electrical Data and Connections

Power Requirements

110 / 220 VAC (jumper selectable)

Frequency

48 to 62 Hz

Power Consumption

max. 7 VA

Signal Output / Load

4 to 20 mA, galvanically isolated

Load Max.

600

pH / ORP input

BNC, screw terminals

Connection terminal

Terminal blocks 5-pole / 17-pole, removable

Mains fuse / fine wire fuse

slow-blow 250 V / 100 mA

EMC Specifications

Emissions

According to EN 50081-1

Susceptibility

According to EN 50082-1

Environmental Conditions (For Indoor Use)

Altitude

Up to 2000 m

Maximum Relative Humidity

80% up to 31

o

C decreasing linearly to 50%

at 40

o

C

Power Supply

Mains supply voltage fluctuations not
exceeding +10% of the nominal voltage

Transient Overvoltages

Category II

Pollution Degree

Category II

Ambient temp. operating range

-10 to +50

o

C

Advertising